Purifies and Distributes High-Quality CLRW Standard Water
The MEDICA R200 from ELGA LabWater is a robust water purification system engineered to meet the high demands of clinical and hospital laboratories. It integrates multiple purification technologies, including reverse osmosis, ultraviolet treatment, and electrodeionization, ensuring consistent delivery of Clinical Laboratory Reagent Water (CLRW) standard water. With its compact design featuring an integral 350-liter reservoir and efficient distribution loop, it seamlessly supports multiple analyzers and laboratory equipment for applications like blood analysis and reagent preparation. The MEDICA R200 operates continuously and is designed for easy automation and integration into existing laboratory workflows. Its efficient purification processes minimize consumable costs and maintenance, offering reliable performance in delivering high-purity water essential for critical clinical diagnostics.
Benefits
- Enhances diagnostic accuracy by providing consistent, high-purity CLRW standard water.
- Reduces operational disruptions with reliable, continuous water supply across multiple analyzers.
- Lowers running costs through energy-efficient purification processes.
- Saves space with an integrated compact design suitable for varied lab setups.
- Simplifies maintenance with long-life consumables and easy system integration.
- Applications
- Clinical laboratories,Hospital laboratories,Glass washers,General laboratory applications,Clinical diagnostics
- End products
- Reagent preparation,Urinalysis panels,Automated chemistry testing,Laboratory equipment maintenance,Pathological glassware cleaning,Blood analysis
- Steps before
- Feedwater Pre-treatment,Water Purification Process
- Steps after
- Distribution to Analyzers,Emergency Bypass Loop,Quality Assurance Testing
- Input ingredients
- potable water,feedwater,inorganic compounds,organic compounds,dissolved gases,microorganisms,bacteria,particulates
- Output ingredients
- CLRW compliant water,purified water,inorganics >10 MΩ.cm,organics <30 ppb,bacteria <10 CFU/ml,particles 0.2 μm filter,silica <0.05 mg/l,delivery flow rate 18-21 l/min
- Market info
- Delivery flow rate
- 18-21 l/min
- Inorganics (resistivity at 25°C)
- >10 MΩ.cm
- Organics (TOC)
- <30 ppb
- Bacteria
- <10 CFU/ml
- Particles
- 0.2 μm filter
- Silica
- <0.05 mg/l
- Reservoir capacity
- 350 liters
- Pump system
- Integrated distribution loop pump
- Reverse Osmosis Module Life Expectancy
- 24-36 months
- Ultra-micron filter Life Expectancy
- 14 months
- Pre-treatment Life Expectancy
- 6 months
- UV Light Life Expectancy
- 12-18 months
- Composite Vent Filter Life Expectancy
- 12 months
- Automation level
- Manual / Automated distribution loop
- Batch vs. continuous operation
- Continuous
- Compact footprint
- Yes
- Water supply method
- CLRW Compliant
- Emergency operation
- Bypass loop for continual supply
- Connection capability
- Feeds multiple analyzers
- Installation flexibility
- Across several laboratories
- Corrosive resistance (e.g. acids)
- Yes
- Biological compatibility
- Suitable for clinical diagnostics
- Cleanability
- High,designed for easy maintenance
- Abrasion resistance
- Yes
- Density/particle size
- N/A
- CLSI Standards
- CLRW Compliant
- Compact footprint
- Integrated 350-liter reservoir
- Control panel type
- Single system design
- Integration possibilities
- Feeds multiple clinical analyzers
- Bypass loop
- For continual water supply